6-4
IM 2560A-01EN
RTD
Type
Selectable Range
Resolution
Pt100
-200.0°C to +850.0°C
0.1°C
RTD USER
-9999.9°C to +9999.9°C
0.1°C
However, within the temperature
range corresponding to resistance
source range (1.00 Ω to 400.00 Ω)
* If you select RTD USER, the resistance corresponding to the user-defined temperature can be
sourced. For a detailed procedure, see section 6.5. For details on definition expressions, see
appendix 6.
Main Setting Display
• The main setting appears flush right at the resolution of each thermocouple or RTD.
• When you change the type or polarity and the specified main setting is outside the selectable range
of the new type, the main setting and polarity displays blink. If you turn any of the main setting dials
in this condition but the main setting still remains outside the selectable range of the new type, the
main setting is set to the maximum or minimum setting of the new type, and the blinking will change
to solid.
<<Command Mnemonic>>
:OUTPut:POLarity[:AUTO] {NORMal|INVerted}
:OUTPut:POLarity:MAXimum {NORMal|INVerted}
:OUTPut:POLarity:MINimum {NORMal|INVerted}
:SOURce:FUNCtion {TCouple|RTD}
:SOURce:LEVel[:AUTO] {<
Temperature
>|MINimum|MAXimum}
:SOURce:LEVel:MAXimum {<
Temperature
>|MINimum|MAXimum}
:SOURce:LEVel:MINimum {<
Temperature
>|MINimum|MAXimum}
:SOURce:SLEVel[:AUTO] {<
Temperature
>|MINimum|MAXimum}
:SOURce:SLEVel:MAXimum {<
Temperature
>|MINimum|MAXimum}
:SOURce:SLEVel:MINimum {<
Temperature
>|MINimum|MAXimum}
6.2 Setting the Temperature