4-2
IM 2560A-01EN
USB
Default gateway
Value:
0.0.0.0
to
255.255.255.255
Example:
0 (GATE 1).0 (GATE 2).0 (GATE 3).0 (GATE 4)
Select the USB communication feature.
GATE
1
GATE
2
GATE
3
GATE
4
MAC address display
Displayed characters: 10 characters, each character ranging
from a to f and 0 to 9
Example:
01.23.45.67.89
MAC
From the previous page
INIT
Options:
TMC
STOR
Remote control → sec. 11.3
Mass storage → sec. 6.5
Options:
CEL
FAR
Celsius (°C)
Fahrenheit (°F)
Displayed unit for thermoelectromotive force output and resistance
output corresponding to the set temperature → sec 6.4
Set the temperature display conditions.
Value:
−1.00 Ω
to
1.00 Ω
(Set the difference between 100.00 Ω and R0.)
Options:
DEG
V.OHM
° (degrees)
V (voltage) or Ω (resistance)
TEMP
MONI
NEXT
Resistance R0 for when the external RJ sensor (Pt100) is at 0°C → sec 6.6
RJC.R0
Temperature unit → sec. 6.7
UNIT
EXEC
Executes clearing
Displays the error number and error code
Displays or clears the error log → sec. 15.5
(Displays log number “0” and error code “NO” if there is no error log.)
ERROR
LOG
NEXT
Moves to/returns from the next menu level
Moves to/returns from the next menu level
Clearing the Error Log
EXEC
Executes initialization
Initializes the settings → sec. 15.8
CLEAR
Firmware version
Logic program version (PCL)
Logic program version (AMP)
Boot program version
Device serial number
Displays the product information. → sec. 15.9
INFO
FIRM
NEXT
Moves to/returns from the next menu level
TEST
EXEC
Executes a self-test
Execute a self-test. → sec. 15.7
PCL
AMP
BOOT
SER.NO
Enter the password.
Performs simple adjustment
CAL
PASS
NEXT
Moves to/returns from the next menu level
For the password entry and simple adjustment procedure, see the following webpage.
http://tmi.yokogawa.com/products/generators-sources/standard/high-voltage-and-high-current-precision-dc-calibrator-2560a/
4.1 SETUP Menu Tree Structure