9-6
IM 2560A-01EN
Setting the GP-IB Address
4.
Turn a
main setting
dial to select ADRS.
The deviation display shows the current GP-IB address.
5.
Turn the
deviation
dials to set the GP-IB address.
• When you change the setting, all PRESET indicators turn on.
• If you decide not to change the setting, step 6 is not necessary.
6.
Flip the
PRESET
switch down.
The setting is confirmed. All the PRESET indicators turn off.
Deviation dials
5. Set the GP-IB address.
PRESET switch
6. Confirm the
settings.
PRESET indicator
Main setting dials (use any)
4. select ADRS.
Setting the Command Type
4.
Turn a
main setting
dial to select TYPE.
The deviation display shows the current command type.
5.
Turn a
deviation dial
to select NORM or COMP.
• When you change the setting, all PRESET indicators turn on.
• If you decide not to change the setting, step 6 is not necessary.
6.
Flip the
PRESET
switch down.
The setting is confirmed. All the PRESET indicators turn off.
Deviation dials (either)
5. Select NORM or COMP.
PRESET switch
6. Confirm the
settings.
PRESET indicator
Main setting dials (use any)
4. Select TYPE.
9.3 Configuring the 2560A GP-IB Settings