pH
91
6.4
Measuring the ORP voltage
The measuring system can measure the voltage of a
solution in conjunction with a ORP electrode, e.g. SenTix
ORP.
Note
Note that, when measuring voltages, it can take a long time
(even several hours) until the measured value becomes
stable.
ORP electrodes are not calibrated. However, you can check
ORP electrodes using a test solution.
1
Call up the
Display
mode with
m
.
2
Call up the measured variable
U
with
46
.
3
Submerse the ORP electrode in the sample.
4
Wait for a stable measured value.