Data storage
206
11.2
Storage of system configurations
Up to 4 different system configurations can be managed
(store, recall, delete and print) by the inoLab Level 3. This
can then be helpful, e.g. if several persons are using the
same measuring system, or different measurement tasks
require different settings.
11.2.1 Saving the system configuration
When storing the system configuration, you store all the
current settings for all probes.
1
Switch to the
Menu
mode with
[Menu]
.
2
Use
46
to select the main
Mem
menu.
3
Use
82
to select the
Settings
menu item.
4
Use
g
to confirm the
Settings
menu item.
5
Select the
Save
menu item with
82
.
6
Confirm the
Save
menu item with
g
.
7
Use
82
to select a free storage location,
Setting-1 ...
-4
.
Free storage locations are displayed in black,
occupied storage locations are gray.
8
Confirm the storing with
g
.
The stored data are displayed.