Technical data
250
Analog output
Automatic switchover between the serial output and analog
recorder output with connection of the cable, AK T-R 2ST.
* The unit can be adjusted in the pH/ION module (see
section 7.2.1).
Accuracy
The following is valid for all recorder
output voltages:
± 0.5% of the corresponding display
value ± 0.5 mV
Internal
resistance
< 5
Ω
(current limit of max. 0.2 mA output
current)
Measuring range
Recorder output
voltage
pH
-2.000 ... +16.000
-0.200 ... 1.600V
U
-999.9 ... +999.9
mV
-1.999 ... +1.999V
-999.9 ... +999.9
mV
-1.999 ... +1.999V
ISE
0 ... 1.999 *
0 ... 19.99 *
0 ... 199.9 *
0 ... 1999 *
0 ... 1.999V
0 ... 1.999V
0 ... 1.999V
0 ... 1.999V
Measuring range
Recorder output
voltage
Oxi
Saturation
0 ... 199.9%
0 ... 600%
0 ... 1.999V
0 ... 0.600V
Concentration
0 ... 20.00 mg/l
0 ... 100.0 mg/l
0 ... 2.000V
0 ... 1.000V
Partial
pressure
0 ... 199.9 mbar
0 ... 1250 mbar
0 ... 1.999V
0 ... 1.250V