Ion concentration
103
Preparatory
activities
Perform the following preparatory activities:
Note
For calibration, use tempered standard solutions with a
temperature that differs by a max. of
±
2 °C from the
temperature of the test sample.
1
Switch on the measuring system with
e
.
2
Connect the electrode to the measuring module.
3
Connect a temperature probe to the measuring
system so that any temperature differences that are
too high will be indicated during calibration.
4
Keep the standard solutions ready.
5
Adjust the temperature of the standard solutions.
6
Switch to the
Menu
mode.
7
Use
46
to select the main
Meas
menu.
8
Confirm the
ISE
menu item with
82
.
9
Confirm the
ISE
menu item with
g
.
The measured variable ISE appears on the display
(
✔
).