Dissolved oxygen (D. O.)
143
Note
In the measuring mode, the system should display an
oxygen saturation of approx. 102% in the air calibration
vessel.
3
Put the D. O. probe into the air calibration vessel.
4
Press the
c
key.
5
Press the
g
key.
h
flashes until a stable value is reached.
The probe symbol shows the relative slope that has
been determined and the probe evaluation (see
page 141).
6
Use
g
to finish the calibration.