![wtw inoLab 740 Скачать руководство пользователя страница 179](http://html1.mh-extra.com/html/wtw/inolab-740/inolab-740_operating-manual_1000561179.webp)
Conductivity
179
10.1.4 Setting the temperature compensation
The calculation of the temperature compensation is based
on the preset reference temperature,
Tref 20
or
Tref 25
(see
).
You can select the following temperature compensations for
the temperature functions:
l
Nonlinear temperature compensation
"nLF"
according to EN 27 888
l
Linear temperature compensation
"lin"
with the
adjustable coefficient of 0.001 to 3.000 %
/
K
l
No temperature compensation "off"
Note
Select the following temperature compensations from the
table according to the respective test sample:
Application notes
Test sample
Temperature
compensation
Display
indicator
Natural water
(ground water,
surface water,
drinking water)
nLF
EN 27 888
n
Ultrapure water
nLF
EN 27 888
n
Other aqueous
solutions
Linear temperature
compensation
Select 0.001 ... 3.000 %/K
q
Salinity
(seawater)
Automatically nLF
according to IOT
k
,
n