Conductivity
186
10.2.2 Setting the cell constant
Note
The cell constant to be set must either be taken from the
operating manual of the measuring cell or is printed on the
measuring cell.
The inoLab 3 measuring system offers the following
possibilities of setting the cell constant:
l
Accepting the cell constant from the last calibration record
(
Cal cell
)
l
Selecting any cell constant within the range 0.250 to
2.500 cm
-1
(
Man cell
)
l
Selecting the fixed cell constant 0.1 cm
-1
(
LR01
)
l
Selecting the fixed cell constant 0.01 cm
-1
(
LR001
)
Note
The last calibration protocol can be displayed at any time
with the aid of the
Cal Prot. ...
function in the
Cal