Conductivity
183
Calibration
procedure
To determine the cell constant, you need a 0.01 mol/l KCl
calibration solution.
1
Start the determination of the cell constant with
c
.
2
Immerse the measuring cell in the prepared
calibration solution.
3
Press the
g
key
in order to start the measurement
with AutoRead to determine the cell constant.
The
h
display indicator flashes until a stable signal
is reached.
The cell constant determined is displayed.
4
Return to the measuring mode with
g
or
m
.