Conductivity
184
Calibration
evaluation
After calibrating, the measuring module automatically
evaluates the calibration. The evaluation appears on the
display.
Calibration record
The measuring module automatically saves the cell
constant in the calibration record. The last calibration
protocol can be displayed at any time with the aid of the
Cal
Prot. function ...
in the
Cal
menu (see section 11.3).
Note
You can automatically print a calibration record after the
calibration. To do so, switch on the printer before calibrating
(print LED lights up green). After a valid calibration, the
record is printed.
Sample printout:
Display
Cell constant [cm
-1
]
0.450 ... 0.500 cm
-1
0.800 ... 1.200 cm
-1
E3
Eliminate the error
according to chapter
14 W
outside the ranges
0.450 ... 0.500 cm
-1
or
0.800 ... 1.200 cm
-1
Conductivity calibration
Cal date: 05.12.1999
Cal time: 22:46
Cal interval: 180 days
Cal standard: 0.01 mol/l KCl
Conductivity/Tref25:
1413 µS/cm
24.6°C
Cell const: 0.987 1/cm
++
User: xxxxx