Dissolved oxygen (D. O.)
146
8.3.2
Drift control (AutoRead)
The AutoRead function checks the stability of the
measurement signal. The stability has a considerable
impact on the reproducibility of the measured values.
When using a probe with integrated stirrer (StirrOx G), the
measurement with AutoRead can be started using the key
on the probe. As soon as a stable measured value is
reached, the stirrer switches itself off.
1
Use
m
to call up the
Display
mode.
2
Call up the oxygen measured variable with
m
.
3
Immerse the D. O. probe in the test sample.
4
Activate the AutoRead function with
a
.
The
h
display indicator appears on the display.
The measured value is frozen (hold function).
5
Start the measurement with AutoRead by pressing
g
or the key on the StirrOx G probe. The
h
display
indicator flashes.
If the StirrOx G probe is used:
3
flashes.
When a stable signal is reached, an acoustic signal is
heard and
h
stops flashing.
If the StirrOx G probe is used: The
3
display
indicator disappears.
6
If necessary, start the next measurement with
AutoRead by pressing
g
or the key on the StirrOx G
probe.
7
To terminate the AutoRead function: Press the
a
key.
Содержание inoLab 740
Страница 18: ...Safety 18...
Страница 27: ...Commissioning 27...
Страница 32: ...Commissioning 32...
Страница 50: ...Basic principles of operation 50...
Страница 92: ...pH 92...
Страница 134: ...Ion concentration 134...
Страница 148: ...Dissolved oxygen D O 148...
Страница 174: ...BOD 174...
Страница 232: ...Data output 232...
Страница 236: ...Maintenance cleaning disposal 236...
Страница 270: ...Lists 270...
Страница 274: ...Lists 274...