Data storage
194
11.1
Storage of measurement data records
Measurement data record memory enables the storage of
measurement data records for subsequent processing or
output.
There are two methods of storing measurement data
records:
l
Manual storage of individual measurement data records
l
Automatic storage of several measurement data records
A measurement data record consists of:
l
Number of the storage location (Mem no.)
l
Date/time
l
Identification (ID)
l
Measured values (e.g. pH and measured temperature
value) with status information
To find data records easily, Mem no., ID and date/time are
available as filter criteria.
The identification (ID), which you allocate yourself, can be
especially useful when doing this (see section 11.1.3).
The ID can consist of a number, or, when using an external
PC keyboard, of digits and letters.
11.1.1 Saving individual measurement data records
You can store individual measurement data records and
assign them an ID that makes it easier to find the data record
again.
1
Switch to the
Display
mode.
2
Start the manual storing with
s
.
The
Store
window shows the current data record.
The instrument automatically allocates a storage
number (Mem no.) to the
Содержание inoLab 740
Страница 18: ...Safety 18...
Страница 27: ...Commissioning 27...
Страница 32: ...Commissioning 32...
Страница 50: ...Basic principles of operation 50...
Страница 92: ...pH 92...
Страница 134: ...Ion concentration 134...
Страница 148: ...Dissolved oxygen D O 148...
Страница 174: ...BOD 174...
Страница 232: ...Data output 232...
Страница 236: ...Maintenance cleaning disposal 236...
Страница 270: ...Lists 270...
Страница 274: ...Lists 274...