pH
80
Calibration interval
The calibration interval determines the time interval between
regular system calibrations.
The terminal reminds you to calibrate regularly by the
flashing probe symbol on the display.
The calibration interval is set to 7 days in the factory. You
can change the interval (1-999 days).
For most applications, we recommend a shorter calibration
interval.
1
Switch to the
Menu
mode.
2
Use
46
to select the main
Cal
menu.
3
Use
82
to select the
Cal Int. ...
menu item.
4
Use
g
to confirm the
Cal Int. ...
menu item.
The
Calibration interval
entry window appears.
5
Option:
Switch between the individual digits with
46
.
6
Increase or reduce the value with
82
.
7
Confirm the value for the calibration interval with
g
.
8
Use
[ESC]
to terminate the
Menu
mode.