Ion concentration
128
Measuring
1
Select the measuring method (see section 7.4.1).
2
Prepare two standard calibration solutions.
3
Perform a two-point calibration according to user
guidance.
4
As soon as a stable value for the second standard
calibration solution is reached, the
display indicator
h
stops flashing. The slope (mV) of
the electrode is displayed after the two-point
calibration. The probe symbol shows the electrode
evaluation after the two-point calibration.
5
Start the measurement with
[Start]
.
The entry for
Standard volume
,
Standard
concentration
and
ISA volume
opens.
6
Enter
Standard volume
,
Standard concentration
and
ISA volume
with
82
and
[Next]
.
7
Thoroughly rinse the electrode with deionized water.
8
Immerse the electrode in the standard solution.