What to do if...
244
14.5
Oxygen system messages
Measuring modules for the inoLab Level 3:
pH, BSB/BOD, pH/Oxi and Multi.
Error message
OFL
Error message
E3
Error message
E7
Cause
Remedy
Display range exceeded
– Not connected
– Connect the probe
– Cable broken
– Replace probe
– Depleted
– Replace probe
– Short-circuit between gold
and lead electrode
– Clean probe and
replace it if necessary
Cause
Remedy
Invalid calibration
– Electrolyte solution
depleted
– Regenerate probe
– Membrane contaminated
– Clean membrane
– Electrode system poisoned
– Regenerate probe
– Worn out
– Replace probe
– broken
– Replace probe
Cause
Remedy
Membrane damaged
– Membrane damaged
– Regenerate probe
– Membrane head not
screwed on tight enough
– Screw membrane head
tight