Conductivity
178
10.1.3 Displaying the conductivity or specific
resistance
You can select the display of the conductivity (in S/cm) or of
the specific resistance (in M
Ω
cm).
1
Switch to the
Menu
mode.
2
Use
46
to select the main
Meas
menu.
3
Select the
Cond setup
menu item with
82
.
4
Confirm the
Cond setup
menu item with
g
.
5
Select the
Cond MOhm
menu item with
82
.
6
Confirm the
Cond MOhm
menu item with
g
.
The conductivity is displayed in the units, M
Ω
cm.
7
Use
[ESC]
to terminate the
Menu
mode.