VM3608A/3616A Programming
CALibration:STORe:AUTO
The Calibration Store auto command allows the new calibration constants, to be saved into the
non-volatile memory, automatically. This command has no effect on the non-volatile memory
when the CAL:SEC:STAT is ON and it generates an error.
CALibration:STORe:AUTO <mode>
Where <mode> is 0 | OFF | 1 | ON. The command will not
execute if calibration security is enabled and, an error is
generated.
EXAMPLES
CALibration:STORe:AUTO ON
Saves the new calibration constants to the non-volatile
memory automatically.
CALibration:STORe:AUTO?
1 (Reports 1, stating that auto store is enabled or ON.)