VTI Instruments Corp.
28
VM3608A/3616A Programming
CALibration:GAIN
The Calibration Gain command is used to set the calibration constant for the gain of the selected
channel and its effect is immediate. The calibration gain command will function only when the
calibration security is disabled, otherwise an error is generated.
CALibration<channel>:GAIN<value>
Where <channel> is 1 through 16 or 1 through 8, referring
to a specific calibration DAC.
Where <value> is -128 to +127.
EXAMPLES
CAL2:GAIN 50
Sets the gain value for Channel 2 to 50.
CAL2:GAIN
?
50 (Returns the gain value for Channel 2, which is currently
set to 50.)