E
XAMPLES OF
SCPI
C
OMMANDS
CALibration:COUNt?
The Calibration Count query returns a number that indicates the number of times the
VM3608A/3616A has been calibrated. The instrument will increment the count every time the
non-volatile memory storing the calibration constants is updated. The non-volatile memory has a
guaranteed minimum of 10,000 cycles.
CALibration:COUNt?
Where the maximum value for count is 16,777,215 after
which it will wrap to 0. There are no query parameters.
EXAMPLES
CAL:COUN?
3 (
Returns a number 3, which indicates the VM3608A/
3616A has been calibrated 3 times.)
CAL:SEC:STAT off,#17VM3616A
Disabling Security
CAL:GAIN 3
Programming Channel 1’s gain
CAL1:ZERO 2
Programming Channel 1’s zero
CAL:STOR
Storing calibration constants to non-volatile memory
CAL:SEC:STAT ON
Enabling security
CAL:COUN?
4 (
Querying calibration count after storing the new
configuration)