VTI Instruments Corp.
104
VM3608A/3616A Command Dictionary
SOURce:SEQuence:DWELl
Purpose
Sets the number of times to loop through a segment in a sequence.
Type
Setting
Command Syntax
[SOURce:]SEQuence:DWEL1 <numeric_value>,<seq_index>
Command Parameters
<numeric_value> = numeric value from 0 to 1048575
<seq_index>
= numeric value from 0 to 4095
*RST Value
All segments set to 1
Query Syntax
[SOURce:]SEQuence:DWEL1? <seq_index>
Query Parameters
<seq_index> = a numeric value from 0 to 4095
Query Response
A numeric value from 0 to 1048575
Description
The Source Sequence Dwell command sets the number of time a sequence will loop
through a specific segment before moving on the next segment. The loop count can range
from 1 to 1048575. If a value of 0 is programmed, the segment will loop indefinitely. A
triggered advance should be setup to advance out of a segment with a loop count of 0.
The sequence index parameter selects which one or more of 4096 possible segments to
apply the command to. Only one channel may be queried at a time.
Examples
Command / Query
Response
(Description)
SEQ:DWEL 10,(@10:25)
SEQ:DWEL? 20
10
Related Commands
N/A