www.vtiinstruments.com
VM3608A/3616A Command Dictionary
107
SOURce:SEQuence:STARt
Purpose
Selects between automatic or triggered starting of a segment in a sequence.
Type
Setting
Command Syntax
[SOURce:]SEQuence:STARt <type>,<seq_index>
Command Parameters
<type>
= AUTOmatic | TRIGgered
<seq_index> = a series of numeric values from 0 to 4095 in standard channel list format
*RST Value
Automatic
Query Syntax
[SOURce:]SEQuence:STARt? <seq_index>
Query Parameters
<seq_index> = Numeric value from 0 to 4095
Query Response
AUTO | TRIG
Description
The Source Sequence Start command selects between automatic or triggered segment
advancement. When a segment has completed its run it will advance to the next segment
in the following ways:
Automatic Start causes a new segment to begin as soon as the previous segment
completes. There is no external trigger required to cause this to happen.
Triggered
Start causes a completed segment to hold its last output value until a trigger
is received to advance to the new segment.
The sequence index parameter selects which one or more of 4096 possible segments to
apply the command to. Only one channel may be queried at a time
Examples
Command / Query
Response
(Description)
SEQ:STAR TRIG,(@1:15)
SEQ:STAR TRIG? 12
TRIG
Related Commands
N/A
Содержание VM3608A
Страница 2: ...VTI Instruments Corp 2 ...
Страница 6: ...VTI Instruments Corp 6 VM3608A 3616A Preface ...
Страница 12: ...VTI Instruments Corp 12 VM3608A 3616A Preface ...
Страница 18: ...VTI Instruments Corp 18 VM3608A 3616A Introduction ...
Страница 134: ...VTI Instruments Corp 134 VM3608A 3616A Command Dictionary ...