VTI Instruments Corp.
132
VM3608A/3616A Command Dictionary
SYSTem:ERRor?
Purpose
Queries the Error Queue
Type
Required SCPI command
Command Syntax
None – Query Only
Command Parameters
N/A
*RST Value
N/A
Query Syntax
SYSTem:ERRor?
Query Parameters
None
Query Response
ASCII string
Description
The System Error query is used to retrieve error messages from the error queue. The error
queue will maintain the two error messages. If additional errors occur, the queue will
overflow and the second and subsequent error messages will be lost. In the case of an
overflow, an overflow message will replace the second error message. See the SCPI
standard Volume 2: Command Reference for details on errors and reporting them.
Examples
Command / Query
Response
(Description)
SYST:ERR?
-350, “Queue overflow”
Related Commands
None