www.vtiinstruments.com
VM3608A/3616A Programming
27
CALibration:DATA
The Calibration Data command is used to send the calibration constants to the VM3608A/3616A
in indefinite or definite length arbitrary block format. The whole data set must be provided,
otherwise an error will be generated. It is important to note that the calibration constants will
change only if the calibration security is off.
CALibration:DATA< block_data>
Where <block_data> is a block of data in IEEE-488.2
definite or indefinite length arbitrary block format. The
calibration constants are implemented as 8-bit values
from -127 to +128.
EXAMPLES
CAL:SEC:STAT OFF,#17VM3616A
Disabling Security
CAL:DATA #232 <block_data>
Sending the data block
Where <block_data> in decimal format would look like:
12300174011021230014367192100156
CAL:SEC:STAT ON
Enabling Security