www.vtiinstruments.com
VM3608A/3616A Command Dictionary
81
CALibration:GAIN
Purpose
Manually sets the calibration constant for the gain of the selected channel.
Type
Setting
Command Syntax
CALibration<channel>:GAIN<value>
Command Parameters
<channel> = 1 - 16 or 1 - 8 referring to a specific Calibration DAC
<value>
= The range for value is -128 to +127
*RST Value
Set to values stored in non-volatile memory.
Query Syntax
CALibration<channel>:GAIN?
Query Parameters
<channel> = 1 - 16 referring to a specific Calibration DAC
Query Response
<value> = -128 to +127 in the specified format
Description
The Calibration Gain command sets the calibration constant for the gain of the selected
channel. Its effect is immediate. If the CALibration:STORe:AUTO ON is set, the
command will save the new constant to the non-volatile memory each time the command
is set. If the CALibration:STORe:AUTO OFF is set, a CALibration:STORe command
must be executed in order to save the new constant. The Calibration Gain command will
function only when the calibration security is disabled, otherwise an error is generated.
The query returns the value from the non-volatile memory rather than the currently used
value and may be different from the constant currently being used.
Examples
Command / Query
Response
(Description)
CAL2:GAIN -120
CAL2:GAIN?
-120
Related Commands
CALibration<channel>:ZERO<value>, CALibration:DATA <block_data>,
CALibration:STORe, CALibration:STORe:AUTO <mode>, FORMat<type>
Содержание VM3608A
Страница 2: ...VTI Instruments Corp 2 ...
Страница 6: ...VTI Instruments Corp 6 VM3608A 3616A Preface ...
Страница 12: ...VTI Instruments Corp 12 VM3608A 3616A Preface ...
Страница 18: ...VTI Instruments Corp 18 VM3608A 3616A Introduction ...
Страница 134: ...VTI Instruments Corp 134 VM3608A 3616A Command Dictionary ...