www.vtiinstruments.com
VM3608A/3616A Introduction
15
D
ATA
M
ODES
Along with static output operation, the VM3608A/3616A provides a FIFO mode where the
selected channels can accept and output a continuous stream of data. The VM3608A/3616A also
offers Arbitrary Waveform Generation (ARB) mode that supports sophisticated looping and
branching to build complex waveforms without the system controller's intervention. The data may
be paced out of the instrument by using either a user-supplied clock or the internal programmable
timer.
C
ALIBRATION
The calibration constants used to correct the data values are stored in non-volatile memory. These
constants are determined when the instrument is calibrated and can be changed as necessary (such
as during routine calibration cycles or when the user selects a new gain setting and wishes to set
the gain accurately). These constants may also be queried at any time via a word serial query and
altered via a word serial command. All calibration is done using calibration DACs to adjust the
gain and offset of each channel. This eliminates the need for removing covers from the unit and
allows for automated calibration.