VM3608A/3616A Command Dictionary
CALibration:STORe
Purpose
Saves the current calibration constants into the non-volatile memory.
Type
Setting
Command Syntax
CALibration:STORe
Command Parameters
None
*RST Value
N/A
Query Syntax
N/A
Query Parameters
N/A
Query Response
N/A
Description
The Calibration Store command saves the current calibration constants into the non-
volatile memory. The CALibration:SECure:STATe must be OFF before using this
command.
Examples
Command / Query
Response
(Description)
CAL:STOR
(Saves the current calibration constants to non-
volatile memory.)
Related Commands
CALibration<channel>:GAIN<value>, CALibration<channel>:ZERO<value>,
CALibration:DATA <block_data>, CALibration:COUNt ?