CALibration:STORe:AUTO
Purpose
Allows the new calibration constants to be saved to non-volatile memory, automatically.
Type
Setting
Command Syntax
CALibration:STORe:AUTO<mode>
Command Parameters
<mode> = 0 | OFF | 1 | ON. The command will not execute if calibration security is
enabled and error -203, “Command protected” is generated.
*RST Value
ON
Query Syntax
CALibration:STORe:AUTO?
Query Parameters
None
Query Response
<mode> = 1 | 0. 1 means the auto store is enabled. 0 means the auto store is disabled.
Description
The Calibration Store Auto command allows the new calibration constants to be saved to
non-volatile memory, automatically. The new calibration constants are automatically
saved to non-volatile memory as they are changed by the CALibration:GAIN and
CALibration:ZERO commands, but not the CALibration:DATA command.
Examples
This example assumes SECURITY is disabled.
Command / Query
Response
(Description)
CAL:STOR:AUTO ON
CAL:STOR:AUTO?
1
Related Commands
CALibration<channel>:GAIN<value>
CALibration<channel>:ZERO<value>