www.vtiinstruments.com
VM3608A/3616A Command Dictionary
77
*WAI
Purpose
Halts execution of additional commands and queries until the No Operation Pending
message is true.
Type
IEEE 488.2 Common Command
Command Syntax
*WAI
Command Parameters
None
*RST Value
N/A
Query Syntax
None
Query Parameters
N/A
Query Response
N/A
Description
The Wait to Continue command halts the execution of commands and queries until the
No Operation Pending message is true. This command makes sure that all previous
commands have been executed before proceeding. It provides a way of synchronizing the
module with its commander.
Examples
Command / Query
Response
(Description)
*WAI
Related Commands
*OPC