www.vtiinstruments.com
VM3608A/3616A Command Dictionary
109
SOURce:VOLTage:FORMat
Purpose
Set the data format for voltage levels
Type
Setting
Command Syntax
SOURce:VOLTage:FORMat <boolean>
Command Parameters
<boolean> = 0 | OFF | 1 | ON
*RST Value
1
Query Syntax
SOURce:VOLTage:FORMat?
Query Parameters
None
Query Response
0 | 1
Description
The Voltage Format command sets the format for the voltage data as implemented in the
word serial command SOURce:VOLTage:DATA, and in Register Access. Note that
Register Access and Word Serial are the opposite of each other. See the
Register Access
section for Register Access details. For Word Serial, when this setting is 0, or OFF,
the binary voltage inputs are
#H0 to #H8000 = -<range> to 0 V
, and
#H8000 to
#HFFFF = 0 to (<range> - LSB)
. For Word Serial, when this setting is 1, or ON, the
binary voltage inputs are
#H0 to #H7FFF = 0 to (<range> - LSB)
, and
#H8000 to
#HFFFF = -<range> to (0 - LSB)
;
where <range> is
10v for the 10 V range
and 20v
for the 20 V range
, and
LSB = (2 * <range>) / 65536
.
See
Programming the DACs via Register Access
Examples
Command / Query
Response
(Description)
SOUR:VOLT:FORM 1
(Sets the binary voltage data format to #H0 to
#H7FFF = positive voltage levels and #H8000
to #HFFFF = negative voltage levels.)
Related Commands
SOURce:VOLTage:DATA
Содержание VM3608A
Страница 2: ...VTI Instruments Corp 2 ...
Страница 6: ...VTI Instruments Corp 6 VM3608A 3616A Preface ...
Страница 12: ...VTI Instruments Corp 12 VM3608A 3616A Preface ...
Страница 18: ...VTI Instruments Corp 18 VM3608A 3616A Introduction ...
Страница 134: ...VTI Instruments Corp 134 VM3608A 3616A Command Dictionary ...