SCAN:LIMit
Purpose
Sets the point in the 512-element scan list array where the scanning should either stop or
loop back to zero.
Type
Setting
Command Syntax
SCAN:LIMit <channel>,<count>
Command Parameters
<channel> = A specific DAC channel of 1 - 16 or 1 - 8
<count>
= A position [ranging from 1 - 512] in the array to stop or loop back
*RST Value
512
Query Syntax
SCAN:LIMit ? <channel>
Query Parameters
<channel> = A specific DAC channel of 1 - 16 or 1 - 8
Query Response
<count> = 1 - 512
Description
The Scan Limit command specifies a position in the 512 element scan list array where the
interrupt routine loading the DACs should either stop or loop back to 0, depending on the
configured scan mode.
Examples
Command / Query
Response
(Description)
SCAN:LIM 2,256
SCAN:LIM? 2
256
Related Commands
SCAN, SCAN:TABle:LOCation <number>,<voltage>, SCAN:TABLe