VM3608A/3616A Programming
31
CALibration:STORe
The Calibration Store command saves the current calibration constants into the non-volatile
memory when the CAL:SEC:STAT is OFF. This command has no effect on the non-volatile
memory when the CAL:SEC:STAT is ON and it generates an error.
CALibration:STORe
The security state is to be off prior to the usage of the
command.
EXAMPLES
CALibration:SECure:STATe OFF,#17VM3640A
Sets the security state off so that the calibration constants can
be stored in the non-volatile memory. Note that the password
is case sensitive.
CALibration:STORe
Saves the current calibration constants to the non-volatile
memory. [Assume that the security is OFF].