VM3608A/3616A Programming
33
CALibration:ZERO
The Calibration Zero command is used to set the calibration constant for the offset of the selected
channel; its effect is immediate. The calibration zero command will function only when the
calibration security is disabled, otherwise an error is generated.
CALibration<channel>:ZERO <value>
Where <channel> is 1 through 16 or 1 through 8 referring
to a specific calibration DAC.
Where <value> is -128 to +127.
EXAMPLES
CALibration 2:ZERO -100
Sets the zero value or offset for Channel 2 to -100.
CALibration 2:ZERO?
-100 (Reports the offset value for Channel 2 that is currently
set as -100.)