VM3608A/3616A Command Dictionary
111
SOURce:VOLTage:RANGe
Purpose
Loads each DAC with a voltage value from the location in its respective “memory lists”
specified by <index>.
Type
Event
Command Syntax
SOURce:VOLTage:RANGe <range>,<channel_list>
Command Parameters
<range>
= 10v | 20v
<channel_list> = Standard format channel list spanning Channels 1 through 16 for the
VM3616A, or Channels 1 through 8 for the VM3608A
*RST Value
20v
Query Syntax
SOURce:VOLTage:RANGe? <channel>
Query Parameters
<channel> = 1 to 16 for the VM3616A or 1 to 8 for the VM3608A.
Query Response
10v | 20v
Description
The Source Voltage Range command selects the output voltage range of the
VM3608A/3616A DAC. In the 10V range, the output spans from +9.99969 V to
-10.00000 V and has a resolution of 305 µV. In the 20 V range, the output spans from
+19.99939 V to -20.00000 V and has a resolution of 610 µV.
Examples
Command / Query
Response
(Description)
SOUR:VOLT:RANG 10V, (@1,2,3)
(Sets the voltage range for Channels 1
through 3 to 10 V.)
Related Commands
N/A