VTI Instruments Corp.
86
VM3608A/3616A Command Dictionary
CALibration:ZERO
Purpose
Used to set the calibration constant for the offset of the selected channel; its effect is
immediate.
Type
Setting
Command Syntax
CALibration<channel>:ZERO<value>
Command Parameters
<channel> = 1 - 16 or 1 - 8 referring to a specific calibration DAC
<value> = –128 to +127.
*RST Value
Set to values stored in the non-volatile memory.
Query Syntax
CALibration<channel>:ZERO?
Query Parameters
<channel> Selects a specific DAC, 1 of 8 for the VM3608A, 1 of 16 for the VM3616A.
Specify 1 to select the first channel.
Query Response
<value> = -128 to +127 in the specified format
Description
The Calibration Zero command is used to set the calibration constant for the offset of the
selected channel; its effect is immediate.
If the CALibration:STORe:AUTO ON is set, the command will save the new constant to
the non-volatile memory each time the command is set. If the
CALibration:STORe:AUTO OFF is set, a CALibration:STORe command must be
executed in order to save the new constant. The CALibration:ZERO command will
function only when calibration security is disabled, otherwise an error is generated.
The query returns the currently used calibration value for the selected channel(s), in the
format specified by the FORMat <type> command.
Examples
Command / Query
Response
(Description)
CAL1:ZERO 115
CAL1:ZERO?
115
FORM HEX
CAL1:ZERO?
#H73
FORM OCT
CAL1:ZERO?
#Q163
FORM:BIN
CAL1:ZERO?
#B1110011
Related Commands
CALibration<channel>:GAIN<value>, CALibration:DATA <block_data>,
CALibration:STORe, CALibration:STORe:AUTO<mode>, FORMat <type>
Содержание VM3608A
Страница 2: ...VTI Instruments Corp 2 ...
Страница 6: ...VTI Instruments Corp 6 VM3608A 3616A Preface ...
Страница 12: ...VTI Instruments Corp 12 VM3608A 3616A Preface ...
Страница 18: ...VTI Instruments Corp 18 VM3608A 3616A Introduction ...
Страница 134: ...VTI Instruments Corp 134 VM3608A 3616A Command Dictionary ...