VM3608A/3616A Programming
29
CALibration:SECure:CODE
The Calibration Secure Code command sets the code required to disable the calibration security.
The calibration security must be disabled in order to change the code string. The default code set
by the factory for VM3616A is ‘VM3616A’ and for VM3608A is ‘VM3608A’.
CALibration:SECure:CODE<code>
Where <code> can be from 1 to 12 ASCII characters in
length entered in IEEE-488.2 definite or indefinite length
arbitrary block format.
EXAMPLES
CALibration:SECure:CODE #17VM3616
Sets the security code for VM3616A in IEEE-488.2 definite or
indefinite length arbitrary block format.
CALibration:SECure:CODE?
#16VM3616 (Returns the calibration security code set as
“VM3616”.