G4 Multi Channel Weighing Instrument
3-35
Range/Alternatives
Explanation and
<default value>
result of alternatives.
Menu ‘EtherNet/IP’
Note: The EtherNet/IP menu is only shown if the EtherNet/IP option is enabled, see
menu ‘Program Options’. See chapter ‘Communication – EtherNet/IP interface’ for
more details on EtherNet/IP configuration and usage.
EtherNet/IP
On
Off
<Off>
Enables/disables the EtherNet/IP interface.
On: EtherNet/IP enabled.
Off: EtherNet/IP disabled.
No of Data Block
Range 0 – 12
< 0 >
Sets the no of Data Blocks that should be mapped to the
memory available over the EtherNet/IP interface. Each Data
Block is configurable.
Block 1 Type
Not In Use
Gross Weight
Net Weight
Disp. Weight
Flow Rate
Input Signal
Level Status
Setp. Status
Input Status
Output Status
Inp./Outp. Status
Lev./Setp. Status
AOUT1-4 Value
AOUT1-2 Value
AOUT3-4 Value
< Not In Use >
Sets the data source for Data Block 1. I.e. sets what data is
transmitted in this Data Block.
Not In Use: The memory area corresponding to this data block
is not updated.
Gross Weight: The gross weight for the selected scale. Floating
point or Integer format.
Net Weight: The net weight for the selected scale. Floating
point or Integer format.
Disp. Weight: The displayed weight (gross or net) for the
selected scale. Floating point or Integer format.
Flow Rate: The Flow Rate for the selected scale. Floating point
or Integer format.
Input Signal: The input signal from the transducer (the mV/V
signal) for the selected scale. Floating point or Integer format.
Level Status: The status of all 32 levels in the instrument.
Floating point format.
Setp. Status: The status of all 16 setpoints in the instrument.
weight for the selected scale. Floating point format.
Input Status: The status of all inputs in the instrument. Floating
point format.
Output Status: The status of all outputs in the instrument.
Floating point format.
Inp./Outp Status: The status of all inputs and all outputs in the
instrument. Integer format.
LEV./Setp Status: The status of all 32 levels and the status of
all 16 setpoints in the instrument. Integer format.
AOUT1-4 Value: The values of analog outputs 1 – 4. Floating
point format.
AOUT1-2 Value: The values of analog outputs 1 – 2. Integer
format.
AOUT3-4 Value: The values of analog outputs 3 – 4. Integer
format.
Note: This parameter is only shown if parameter
‘No Of Data Block‘ is set to 1 or greater.
Содержание G4
Страница 2: ......
Страница 18: ...Technical Manual 1 14 ...
Страница 39: ...G4 Multi Channel Weighing Instrument 3 7 ...
Страница 104: ...Technical Manual 5 18 ...
Страница 158: ...Technical Manual 6 54 ...
Страница 178: ...Technical Manual 7 20 ...
Страница 186: ...Technical Manual 8 8 ...
Страница 196: ......
Страница 197: ...G4 Multi Channel Weighing Instrument Appendix 1 ...
Страница 198: ...Technical Manual ...
Страница 199: ......