Technical Manual
3-18
Range/Alternatives
Explanation and
<default value>
result of alternatives.
1:No Motion Dely
Range:
0 to 10.0
Unit: s
<1.0>
Delay in seconds from detection of stable weight
until the Motion status goes off.
1:Motion Check
Off
On
<Off>
Off: Only zero adjustment is inhibited during motion.
On: Inhibits zero adjustment, taring, and printing during motion.
1:Min. Weight Pr
Range:
0 to 999999
Unit:
Measurem. unit
<0.0>
Defines the smallest allowed weight value that to be printed.
0 gives no restrictions.
1:Overload Check
Off
Unipolar
Bipolar
<Off>
Check of overload is performed according to this set-up.
Off: No check is performed. Weight will be reported
up to the limits of the AD converter.
Unipolar: Overload status will be set if ‘Overload limit’ is
exceeded. Underload status will be set if the gross weight is
below minus (-)9 * Resolution.
Bipolar: Overload status will be set if Overload limit
is exceeded.
Underload status will be set if the gross weight is below
minus (-) Overload limit.
1:Overload Limit
Range:
0.5 to 999999
Unit:
Measurem. unit
<Ca
9 * Resolution>
Should be set to the max gross weight that the weight indicator
is allowed to report. This parameter is always set to default
value when changing Resolution or Capacity. If ‘Bipolar’
overload check is selected the weight will be reported up to the
Overload limit and down to the minus (-) Overload limit.
Note: this parameter is only shown if parameter
‘1: Overload Check’ is set to ‘Unipolar’ or ‘Bipolar’.
Содержание G4
Страница 2: ......
Страница 18: ...Technical Manual 1 14 ...
Страница 39: ...G4 Multi Channel Weighing Instrument 3 7 ...
Страница 104: ...Technical Manual 5 18 ...
Страница 158: ...Technical Manual 6 54 ...
Страница 178: ...Technical Manual 7 20 ...
Страница 186: ...Technical Manual 8 8 ...
Страница 196: ......
Страница 197: ...G4 Multi Channel Weighing Instrument Appendix 1 ...
Страница 198: ...Technical Manual ...
Страница 199: ......