Deadweight calibration related parameters
Used when the scale is calibrated with weights. The instrument automatically reads the
corresponding transducer signals.
1:No of Cal. P
Range:
2 to 6
<2>
Number of calibration points.
1:Value Cal. P1
Range: +/–999999
Unit:
Measurem. unit
<0>
This parameter defines the load on the scale in the lowest
calibration point, normally 0.
The live gross weight value will be displayed when the enter
button is pressed. Press enter button a second time to edit the
calibration point value.
1:Value Cal. P2
Range: +/–999999
Unit:
Measurem. unit
<500>
This parameter defines the load on the scale in the second
calibration point.
1:Value Cal. P3
Range: +/–999999
Unit:
Measurem. unit
<0>
This parameter defines the load on the scale in the third
calibration point.
1:Value Cal. P4
Range: +/–999999
Unit:
Measurem. unit
<0>
This parameter defines the load on the scale in the fourth
calibration point.
1:Value Cal. P5
Range: +/–999999
Unit:
Measurem. unit
<0>
This parameter defines the load on the scale in the fifth
calibration point.
1:Value Cal. P6
Range: +/–999999
Unit:
Measurem. unit
<0>
This parameter defines the load on the scale in the sixth
calibration point.
Содержание G4
Страница 2: ......
Страница 18: ...Technical Manual 1 14 ...
Страница 39: ...G4 Multi Channel Weighing Instrument 3 7 ...
Страница 104: ...Technical Manual 5 18 ...
Страница 158: ...Technical Manual 6 54 ...
Страница 178: ...Technical Manual 7 20 ...
Страница 186: ...Technical Manual 8 8 ...
Страница 196: ......
Страница 197: ...G4 Multi Channel Weighing Instrument Appendix 1 ...
Страница 198: ...Technical Manual ...
Страница 199: ......