Technical Manual
3-26
Range/Alternatives
Explanation and
<default value>
result of alternatives.
Table calibration related parameters
Used when the scale is calibrated with recorded values from a previous calibration,
normally a deadweight calibration.
1:No of Cal. P
Range:
2 to 6
<2>
Number of calibration points.
1:Value Cal. P1
Range: +/–999999
Unit:
Measurem. unit
<0>
In this parameter, enter the recorded value for the load
in the first calibration point.
The live gross weight value will be displayed when the
enter button is pressed. Press enter button a second
time to edit the calibration point value.
1:Value Cal. P2
Range: +/–999999
Unit:
Measurem. unit
<500>
In this parameter, enter the recorded value for the load
in the second calibration point.
1:Value Cal. P3
Range: +/–999999
Unit:
Measurem. unit
<0>
In this parameter, enter the recorded value for the load
in the third calibration point.
1:Value Cal. P4
Range: +/–999999
Unit:
Measurem. unit
<0>
In this parameter, enter the recorded value for the load
in the fourth calibration point.
1:Value Cal. P5
Range: +/–999999
Unit:
Measurem. unit
<0>
In this parameter, enter the recorded value for the load
in the fifth calibration point.
1:Value Cal. P6
Range: +/–999999
Unit:
Measurem. unit
<0>
In this parameter, enter the recorded value for the load
in the sixth calibration point.
Содержание G4
Страница 2: ......
Страница 18: ...Technical Manual 1 14 ...
Страница 39: ...G4 Multi Channel Weighing Instrument 3 7 ...
Страница 104: ...Technical Manual 5 18 ...
Страница 158: ...Technical Manual 6 54 ...
Страница 178: ...Technical Manual 7 20 ...
Страница 186: ...Technical Manual 8 8 ...
Страница 196: ......
Страница 197: ...G4 Multi Channel Weighing Instrument Appendix 1 ...
Страница 198: ...Technical Manual ...
Страница 199: ......