G4 Multi Channel Weighing Instrument
9-5
Error
code
Explanation
111
Taring not allowed (negative gross weight).
Taring is not allowed at negative gross weight if parameter ‘Overload check’
is set to Unipolar.
112
Instrument in net mode.
Zero setting requires that the instrument is in gross mode. However, if you
try to transmit a zero setting command while the scale is in net mode you’ll
receive this error code.
113
Outside zero setting limits.
Adjustment of the zero setting during operation is only possible if
the accumulated correction required is within -1% and +3% of
the capacity set-up.
Consequently, if you transmit a zero setting command while the required
adjustment is outside allowed range you will receive this error code.
114
Setpoint(s) not in use.
Activation/Deactivation of setpoint(s) can not be done, as at least one of
the setpoint sources is/are set to ‘not in use’.
115
Weight below min weight for printing
The weight is below the lowest allowed for printing. The low limit is set by
parameter ‘Min. Weight Print’. See chapter ‘Set-up’.
116
Net mode not allowed.
Net weight cannot be shown when the tare value equals 0.
117
Flow display not allowed.
Flow rate is not enabled for the actual scale.
118
Zero setting not allowed.
Zero setting is not allowed when the calibration menu is displayed
for the actual scale.
122
Printer buffer overflow.
Communication print command error response. Given when the printer
buffer for a used port is full. The current communication set-up and possibly
the printer cannot handle the amount of print commands that been issued
(during a limited time) over communication, print button and digital inputs
totally.
123
Printing error.
Communication print command error response. There is an undefined
internal error regarding printer handling. Try the command again or restart
the instrument. Contact your supplier if the error persists.
124
Program option not enabled
A command that applies to a program option that is not enabled, was given.
The command is not executed.
125
Wrong batching state
A batching command was given while the scale was in a batching state not
allowing this command. The command was not executed.
126
Not allowed during batching
The parameter or Preset Tare that the user tried to edit is not allowed to be
change during batching.
Содержание G4
Страница 2: ......
Страница 18: ...Technical Manual 1 14 ...
Страница 39: ...G4 Multi Channel Weighing Instrument 3 7 ...
Страница 104: ...Technical Manual 5 18 ...
Страница 158: ...Technical Manual 6 54 ...
Страница 178: ...Technical Manual 7 20 ...
Страница 186: ...Technical Manual 8 8 ...
Страница 196: ......
Страница 197: ...G4 Multi Channel Weighing Instrument Appendix 1 ...
Страница 198: ...Technical Manual ...
Страница 199: ......