G4 Multi Channel Weighing Instrument
3-21
Range/Alternatives
Explanation and
<default value>
result of alternatives.
1:Auto Deriv. T
Off
On
<On>
The derivation time can be entered manually or calculated
automatically by the instrument.
Off: Manual entering of derivation time.
On: Automatic calculation of derivation time.
Note: this parameter is only shown if parameter ‘1:Flow
calculation’ is set to ‘On’.
1:Deriv. Time
Range:
0.10 – 3600.00
Unit: s
The flow rate is the weight change over the latest elapsed
”Derivation time” divided by this time.
At calibration the derivation time is always adjusted
by the instrument to the nearest possible value (depending on
the filter time).
Note: this parameter is only shown if parameter ‘1:Flow
calculation’ is set to ‘On’.
Note: This parameter is read only if
‘1:Auto derivation time’ is set to ‘On’.