G4 Multi Channel Weighing Instrument
2-1
2. Installation
Mechanical installation
See chapter ‘Introduction – Technical data’ for references to RM mechanical measures.
G4, RM type, should be used in a dry, clean environment or it must be mounted
in a cabinet, protecting from water, dust etc. The instrument is prepared for
mounting on a flat surface by DIN rail according to Technical data.
The instrument needs minimum 200 mm space above this surface.
To remove the instrument from the rail, the black lever at the lower bottom side of
the instrument must be engaged.
Each instrument should have a free space of minimum 30 mm to the left
and to right, 100 mm above and below the instrument.
Cable ducts may be mounted in this free space.
All electrical connections to the instrument modules are made to terminals
at the front side so enough room for these terminals should be arranged.
CABLE DUCT
LOCK LEVER
TERMINALS
100
100
200
RM instrument.
Recommended free space at the instrument.