Technical Manual
6-50
Examples
Example 1: Setting ‘Level 1 Value’ to 123.5 (Writing to float value register).
1. Make sure that the previous command was not 251.
Set command byte (00) to 0 if previous command was 251.
2. Set number of registers (2) to write in byte 01.
3. Set start address (46026) in bytes 02 and 03.
See chapter ‘Communication – Modbus protocol – General registers’.
4. Set the following four bytes (04 – 07) to the value to write (123.5).
5. Enter the command (251) in byte 00.
Byte
Description
Dec
Hex
00
Command
251
FB
01
Number of registers to write
2
02
02 – 03
Start address
46026
B3 CA
04 – 07
Value to write in register 1 and 2
123.5
42 F7 00 00
08 – 15
Write registers 3 – 6.
Example 2: Setting ‘Setpoint 14 value’ to 123.5 (Writing to integer value register).
1. Make sure that the previous command was not 251.
Set command byte (00) to 0 if previous command was 251.
2. Set number of registers (3) to write in byte 01.
3. Set start address (42169) in bytes 02 and 03.
See chapter ‘Communication – Modbus protocol – General registers’.
4. Set the following four bytes (04 – 07) to the value to write 123.50,
i.e. 12350 and two decimals.
5. Set bytes 08 and 09 to number of decimals. In this example is 2 decimals used.
6. Enter the command (251) in byte 00.
Byte
Description
Dec
Hex
00
Command
251
FB
01
Number of registers to write
3
03
02 – 03
Start address
42169
A4 B9
04 – 07
Value to write in register 1 and 2
12350
00 00 30 3E
08 – 09
Value to write to register 3
2
00 02
10 – 15
Write registers 4 – 6.
Содержание G4
Страница 2: ......
Страница 18: ...Technical Manual 1 14 ...
Страница 39: ...G4 Multi Channel Weighing Instrument 3 7 ...
Страница 104: ...Technical Manual 5 18 ...
Страница 158: ...Technical Manual 6 54 ...
Страница 178: ...Technical Manual 7 20 ...
Страница 186: ...Technical Manual 8 8 ...
Страница 196: ......
Страница 197: ...G4 Multi Channel Weighing Instrument Appendix 1 ...
Страница 198: ...Technical Manual ...
Страница 199: ......