Technical Manual
6-18
Instrument state
This register contains the state of the G4 instrument unit.
Code
Description
00
‘Starting up’ state.
The instrument is starting up after a reset or power on.
01
‘Wait for start’ state.
The instrument is waiting for a start command to go in process.
02
‘Warming up state’
The parameter ‘Warm up time’ is set to a value other than zero, and
the instrument is waiting for the warming up time to pass.
03
‘Normal’ state.
There are no parameter errors in the system.
Note:
Weight errors still indicates normal state.
04
‘Error’ state.
An error has been detected during start up of the instrument.
05
‘Fatal error’ state.
An error has been detected during start up of the instrument.
It’s not possible to enter any other state from here.
06
‘Power fail’ state.
A power fail has been detected and the instrument will respond to
communication for a short moment until the power supply reserves are empty
(this is normally just a few ms). Weight values should be regarded as invalid.
Instrument error
This register holds the overall error code for the instrument.
Normally this register should contain ‘00’ which means no error.
Содержание G4
Страница 2: ......
Страница 18: ...Technical Manual 1 14 ...
Страница 39: ...G4 Multi Channel Weighing Instrument 3 7 ...
Страница 104: ...Technical Manual 5 18 ...
Страница 158: ...Technical Manual 6 54 ...
Страница 178: ...Technical Manual 7 20 ...
Страница 186: ...Technical Manual 8 8 ...
Страница 196: ......
Страница 197: ...G4 Multi Channel Weighing Instrument Appendix 1 ...
Страница 198: ...Technical Manual ...
Страница 199: ......