<5. Startup>
5-5
IM 11Y01D01-01EN
The procedure for changing the TDLS8000 IP address and reconnecting is as follows. If you
change the TDLS8000 subnet mask, you will also need to change the YH8000 subnet mask as
described in “Setting the YH8000 IP address” on the next page before reconnecting.
(1) Tap in the lower right of the screen.
The tag names (or the serial numbers if not assigned) of the connected TDLS8000s appear.
Select the TDLS8000 that you want to change the IP address of.
In the example in the figure, the tag name is “TAG1.” Tag names are not assigned to
TDLS8000s with factory default settings, so serial numbers are displayed instead.
After selecting, tap OK. A password input screen will appear.
(2) Enter the password and tap Enter. A configuration screen will appear. The factory default
password is “1234.”
From the menu, select Configuration>>System>>Communication>>TCP/IP. A warning
screen explaining that the instrument will restart if the IP address is changed appears.*1 Tap
OK.
*1:
The analyzer will restart after changing IP settings”
3rd Edition: Jun. 10, 2016-00
Содержание TDLS8000
Страница 1: ...User s Manual TDLS8000 Tunable Diode Laser Spectrometer IM 11Y01D01 01EN IM 11Y01D01 01EN 3rd Edition...
Страница 10: ...ix IM 11Y01D01 01EN LT LV PL EST SLO H BG RO M CZ SK 3rd Edition Jun 10 2016 00...
Страница 25: ...Blank Page...
Страница 45: ...Blank Page...
Страница 115: ...Blank Page...
Страница 151: ...Blank Page...
Страница 231: ...Blank Page...
Страница 239: ...Blank Page...
Страница 243: ...Blank Page...
Страница 261: ...Blank Page...
Страница 303: ......
Страница 305: ......
Страница 307: ......
Страница 309: ......
Страница 311: ......
Страница 313: ...Blank Page...