Specifications
2.1
TDLS8000 Tunable Diode Laser
Spectrometer Specifications
Measurement object: O
2
, CO, CO or CH
4
, CO
2
, CO + CO
2
, H
2
O, NH
3
, NH
3
+ H
2
O, H
2
S, HCl
concentration in combustion exhaust gas and process gas
If other gas measurements are required, consult with Yokogawa.
Measurement system: Tunable diode laser spectroscopy
Light source; Near-infrared tunable diode laser
Measured components and ranges:
Measured component
Min. range
Max. range
O
2
0-1%
0-25%
CO(ppm)
0-200 ppm
0-10000 ppm
CO or CH
4
(*3)
CO
0-200 ppm
0-10000 ppm
CH
4
0-5%
NH
3
0-30 ppm
0-5000 ppm
H
2
O(ppm) in non HC (*1)
0-30 ppm
0-30000 ppm
H
2
O(ppm) in HC (*2)
0-30 ppm
0-30000 ppm
CO (%)
0-20%
0-50%
CO (%) + CO
2
(%)
0-30%
0-100%
NH
3
+ H
2
O
NH
3
0-30 ppm
0-5,000 ppm
H
2
O
0-5%
0-50%
H2S
0-5%
0-100%
CO
2
(%) High Range
0-1%
0-5%
CO
2
(%) Extend. Range
0-30%
0-50%
H
2
O (%)
0-10%
0-100%
HCl
0-50 ppm
0-5,000 ppm
*1:
Non hydrocarbon background
*2:
Hydrocarbon background
*3:
Please consult with Yokogawa if CO or CH
4
ingredient coexists.
Please consult with Yokogawa if the measuring range for your sample gas is outside of the above
ranges
Optical path length:
Optical distance between the laser unit and the sensor control unit
Standard; 0.5 to 6 m (Application dependent)
Max;
30 m (With optional Large Aperture Optics (LAO))
Note: If your optical path length is under 0.5 m or over 30 m, please consult with Yokogawa.
Safety and EMC conforming standards:
Safety Conforming Standards:
CE
EN61010-1, EN61010-2-030
UL
UL61010-1, UL 61010-2-030
CSA
CAN/CSA-C22.2 No.61010-1, CAN/CSA-C22.2 No.61010-2-030
GB
GB30439 Part 1
Installation altitude: 2000 m or less
Installation category: I (Anticipated transient overvoltage 330V)
Measuring category: O (Other)
Pollution degree:
2, Indoor/Outdoor use
Note: Installation category, called overvoltage category, specifies impulse withstand voltage. Pollution degree indicates the degree of
existence of solid, liquid, gas or other inclusions which may reduce dielectric strength.
EMC Conforming Standards:
CE
EN55011 Class A Group 1
EN61326-1 Class A Table 2 (For use in industrial location), EN61326-2-3
RCM
EN55011 Class A Group 1
KC
KN11 Class A Group 1, KN61000-6-2 (Korea Electromagnetic Conformity)
3rd Edition: Jun. 10, 2016-00
Содержание TDLS8000
Страница 1: ...User s Manual TDLS8000 Tunable Diode Laser Spectrometer IM 11Y01D01 01EN IM 11Y01D01 01EN 3rd Edition...
Страница 10: ...ix IM 11Y01D01 01EN LT LV PL EST SLO H BG RO M CZ SK 3rd Edition Jun 10 2016 00...
Страница 25: ...Blank Page...
Страница 45: ...Blank Page...
Страница 115: ...Blank Page...
Страница 151: ...Blank Page...
Страница 231: ...Blank Page...
Страница 239: ...Blank Page...
Страница 243: ...Blank Page...
Страница 261: ...Blank Page...
Страница 303: ......
Страница 305: ......
Страница 307: ......
Страница 309: ......
Страница 311: ......
Страница 313: ...Blank Page...