<Appendix 8 Explosion Protected Type Instrument>
App.8-16
IM 11Y01D01-01EN
(2) Name Plate
MODEL:
Specified model code
AMB.TEMP:
Specified ambient temperature range
SUFFIX:
Specified suffix code
OUTPUT:
Specified analog output range
NO.:
Serial number
STYLE:
Specified style code
SUPPLY:
Specified supply voltage and wattage
MANUFACTURED: Month and year of production
Ex marking:
Ex d IIC T5
CL II/III, DIV 1, GPS EFG
TYPE 4X, IP66
DUAL SEAL
Main Name Plate (Sensor Control Unit side)
Sub Name Plate (Laser Unit side)
(3) Installation
• The installation of the equipment should be carried out by the engineers or other
professional personnel of the related expertise. The installation should not be carried out by
operators or other unprofessional personnel.
• All wiring shall comply with Canadian Electrical Code Part I and Local Electrical Codes.
• In hazardous location, wiring shall be in conduit.
WARNING
AVERTISSEMENT
A SEAL SHALL BE INSTALLED WITHIN 50 mm OF THE ENCLOSURE.
PLACER UN MATÉRIAU ÉTANCHE A 50 mm AUTOUR DU BOÎTIER.
3rd Edition: Jun. 10, 2016-00
Содержание TDLS8000
Страница 1: ...User s Manual TDLS8000 Tunable Diode Laser Spectrometer IM 11Y01D01 01EN IM 11Y01D01 01EN 3rd Edition...
Страница 10: ...ix IM 11Y01D01 01EN LT LV PL EST SLO H BG RO M CZ SK 3rd Edition Jun 10 2016 00...
Страница 25: ...Blank Page...
Страница 45: ...Blank Page...
Страница 115: ...Blank Page...
Страница 151: ...Blank Page...
Страница 231: ...Blank Page...
Страница 239: ...Blank Page...
Страница 243: ...Blank Page...
Страница 261: ...Blank Page...