<Appendix 8 Explosion Protected Type Instrument>
App.8-17
IM 11Y01D01-01EN
50 mm
50 mm
Note: Laser unit shall have conduit and sealing fitting in the same manner as Sensor control unit.
• Unused apertures shall be closed with suitable flameproof certified blanking elements. The
plugs that are installed in the instrument are certified as a part of the instrument.
• In case of ANSI 1/2 and 3/4 plugs, ANSI hexagonal wrench should be applied to screw in.
• Mounting orientation of TDLS8000 is within 30 degrees relative to the horizontal direction.
• If the TDLS8000 is mounted in an area where the use of Class II and III equipment is
required, it shall be installed in such a way that the risk from electrostatic discharges and
propagating burst discharges caused by rapid flow of dust is avoided.
• A mark indicating the electrical connection type is stamped near the electrical connection
port. These marks are as followed.
M
Location of the marking
3rd Edition: Jun. 10, 2016-00
Screw size
Marking
ISO M20x1.5 female
M25x1.5 female
M
ANSI 1/2 NPT Female
3/4 NPT female
N
Note: This type is only NPT thread available.
(4) Operation
• Note a warning label worded as follows.
WARNING
AVERTISSEMENT
• DO NOT OPEN WHEN AN EXPLOSIVE ATMOSPHERE MAY BE PRESENT.
• POTENTIAL ELECTROSTATIC CHARGING HAZARD.
• FOR INSTALLATION AND SAFE USE, READ IM 11Y01D01-01.
• A SEAL SHALL BE INSTALLED WITHIN 50 mm OF THE ENCLOSURE.
• NE PAS OUVRIR NE ATMOSOHERE EXPLOSIVE.
• DANGER : ELECTRICITÉ STATIQUE
• LISEZ IM 11Y01D01-01 POUR INSTALLATION ET SÉCURITÉ.
• PLACER UN MATÉRIAU ÉTANCHE A 50 mm AUTOUR DU BOÎTIER.
• Take care not to generate mechanical spark when access to the instrument and peripheral
devices in hazardous locations.
• Electrostatic charge may cause an explosion hazard. Avoid any actions that cause the
generation of electrostatic charge, such as rubbing with dry cloth on coating face of product.
• Purge inside of the each enclosure using nitrogen for at least 60 minutes at 10 kPa before
turning on the power to prevent from pressurizing internal gas by purge gas. After stoppage
of purge gas, explosive gas may be inside of the enclosures.
Содержание TDLS8000
Страница 1: ...User s Manual TDLS8000 Tunable Diode Laser Spectrometer IM 11Y01D01 01EN IM 11Y01D01 01EN 3rd Edition...
Страница 10: ...ix IM 11Y01D01 01EN LT LV PL EST SLO H BG RO M CZ SK 3rd Edition Jun 10 2016 00...
Страница 25: ...Blank Page...
Страница 45: ...Blank Page...
Страница 115: ...Blank Page...
Страница 151: ...Blank Page...
Страница 231: ...Blank Page...
Страница 239: ...Blank Page...
Страница 243: ...Blank Page...
Страница 261: ...Blank Page...
Страница 303: ......
Страница 305: ......
Страница 307: ......
Страница 309: ......
Страница 311: ......
Страница 313: ...Blank Page...